FLUICAL®
Bench-Top Calibration System
General
Bronkhorst High-Tech have developed a bench-top calibration system to
meet the increasing demands of users of measurement and control equipment
within the ISO-9000 quality control system. The FLUICAL® Bench-Top
Calibrator offers great accuracy, traceable to national standards. As an
option, our ISO 17025 accredited Bronkhorst
Calibration Centre can offer calibration services for FLUICAL flow
calibration systems.
Description
With Bronkhorst's FLUICAL® Bench-Top Calibration System (Mass)
Flow Meters and Controllers for gases can be accurately calibrated. It
is an automatic system that consists of one or more calibration devices
and an interface / readout / control unit. For flows to a maximum of 10
l/min piston provers can be used, whilst higher flow rates can be calibrated
using high-accuracy rotor meters (maximum 100 m3/h).
The piston prover is a volumetric calibration device consisting of a
precision bore borosilicate glass tube and a mercury sealed piston. When
a gas flow enters into the vertical mounted glass tube the piston will
move upwards. A number of sensors have been mounted along the wall of the
tube to detect the presence of the piston. The volume between these sensors
has been calibrated and is therefore a fixed known volume. Together with
the travel time, pressure and temperature in the glass tube, the flow at
reference conditions can be calculated.
The FLUICAL® piston provers and rotor meters can be controlled
by means of the Bronkhorst High-Tech series E-7000 digital readout and
control system. With an optional RS232 - interface the FLUICAL®
system can be connected to and controlled be means of a personal computer.
Normally the FLUICAL® calibration devices are provided with
Bronkhorst High-Tech calibration certificates. If required the calibration
devices can be certified by VSL, the independent measurement authority
in the Netherlands.
Capacities
| Range up to |
Calibration device |
| 1000 mln/min Air |
Model T-916-TD (Piston prover) |
| 10 ln/min Air |
Model T-950-TD (Piston prover) |
| 25 m3n/h Air |
Model R-025-TD (Rotor meter) |
| 65 m3n/h Air |
Model R-065-TD (Rotor meter) |
| 100 m3n/h Air |
Model R-100-TD (Rotor meter) |
Uncertainty 0,2 ... 0,4 RD (2S)
|