Bronkhorst Calibration Centre
ISO/IEC 17025:2005 accredited
| The Bronkhorst Calibration Centre is accredited ISO/IEC 17025:2005
by the Dutch Accreditation Council (RvA) under registration number K 127.
Herewith is certified that our laboratory’s Quality Assurance system, calibration
methods and technical competences have been inspected and audited and meet
all necessary requirements. |
Gas flow meter / controller calibrations and adjustments
All Bronkhorst instruments are factory calibrated using highly accurate calibration devices, traceable to (inter)national standards. In addition, accredited calibrations may be required for various reasons, e.g. to meet specific demands from authorities, or for instruments that will be used as reference standards. With more than 30 years of calibration experience Bronkhorst decided to establish a calibration laboratory and apply for accreditation in the field of mass flow and volume flow gas calibration.
Upon accreditation in 2010, calibration services, including adjustments, can be offered in the range of 0.7 mln/min up to 6200 ln/min for both new and existing Bronkhorst gas flow meters / controllers and gas flow calibration devices. In addition the calibration centre can also perform gas flow calibrations for volumetric type of flow meters and other brands of mass flow meters/controllers.
Benefits
- Enhanced confidence of customers and authorities
- Cost effective and avoiding expensive retesting
- “One-stop service” for purchasing flow measuring equipment with accredited calibration
- Qualified staff with quality first approach and an eye for customers needs
- Excellent measurement capabilities
- Actual gas calibrations
Scope for Mass flow calibration / Volume flow calibration
| Gas MASS flow | Gas VOLUME flow | Gas MASS flow | Gas VOLUME flow | CMC 1) |
| (Air) | (Air) | (N2, Ar, CO2, He, H2) | (N2, Ar, CO2, He, H2) | (k=2) |
| (at 21 ± 2°C and 50 ± 20% RH) |
(at 21 ± 2°C and 50 ± 20% RH) |
(at 21 ± 2°C and 50 ± 20% RH) |
(at 21 ± 2°C and 50 ± 20% RH) |
(% of Reading) |
| 0.66–3.3 mln/min2) | 0.71–3.5 ml/min2) | 0.66–3.3 mln/min4) | 0.71–3.5 ml/min4) | 0.43% |
| 3.3–15.5 mln/min2) | 3.5–16.6 ml/min2) | 3.3–15.5 mln/min4) | 3.5–16.6 ml/min4) | 0.31% |
| 0.0155–25 ln/min2) | 0.0166–26.8 l/min2) | 0.0155–25 ln/min4) | 0.0166–26.7 l/min4) | 0.25% |
| 5.04–1008 ln/min2) | 5.41–1082 l/min2) |
- |
- |
0.21% |
|
1008–6202 ln/min3) |
1082–6645 l/min3) |
- |
- |
0.22% |
2) at 1…8 bar a
3) at 1…7 bar a
4) at 1…10 bar a
Click here to review the scope at the website of the Dutch Accreditation Council (Raad voor Accreditatie- RvA)
Brochures
For more info see brochure Bronkhorst Calibration Centre
English
中文
Other languages...
