Bronkhorst

μ-FLOW L01V12

Ultra Low-Flow Thermal Liquid Mass Flow Controller

  • Fast and accurate measuring signal
  • Insensitive to mounting position
  • Very small internal volume
  • Suitable for liquids with low boiling points
  • Analog, RS232 or fieldbus communication
Get a quote Request a demo Support
μ-FLOWL01V12

Liquid Mass Flow Controllers for low flow rates

Bronkhorst® model L01V12 Liquid Flow Controllers (LFCs) are suited for precise measurement and control of flow ranges between 5…100 mg/h and 0,1...2 g/h at operating pressures up to 100 bar. The LFC consists of a thermal mass flow sensor and a microprocessor based pc-board with signal and fieldbus conversion and a PID controller for mass flow control by means of an integrated control valve.

μ-FLOW series are equipped with a digital pc-board, offering high accuracy, excellent temperature stability and fast response. The main digital pc-board contains all of the general functions needed for measurement and control. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide DeviceNet™, PROFIBUS DP, Modbus or FLOW-BUS protocols.


Technical specifications

Measurement/Control system
Accuracy (incl. linearity)
(based on actual calibration)
±2% FS
Turndown 1:20 (5...100%)
Repeatability < 0,2% FS (typical H2O)
Settling time 2...4 seconds
Max. Kv-value 2,37x10-3
Max. fluid viscosity 0.1 Pa*s
Operating temperature 5…50°C
Temperature sensitivity ±0,2% FS/°C
Attitude sensitivity negligible
Warm-up time approx.10 min. for accuracy ± 2% FS
Mechanical parts
Material (wetted parts) stainless steel 316L/320;
others on request
Pressure rating 100 bar abs
Max. ΔP 10 bar dif.
Process connections 1/16” or 1/8” OD compression type; other on request
(<1 g/h we advise to use 1/16" only)
Purge connection 1/16” OD compression type
Seals Kalrez®-6375;
others on request
Ingress protection (housing) IP40
Electrical properties
Power supply +15…24 Vdc +/-10%
Max. power consumption Supply
15 V
24 V
at voltage I/O
285 mA
250 mA
at current I/O
305 mA
270 mA
PROFIBUS DP add 53 mA (15 V supply) or 30 mA (24 V supply)
DeviceNet™ add 48 mA (24 V supply)
Analog output/command 0...5 (10) Vdc or 0 (4)...20 mA (sourcing output)
Digital communication standard: RS232;
options: PROFIBUS DP, DeviceNet™, Modbus RTU or ASCII, FLOW-BUS
Electrical connection
Analog/RS232 9-pin D-connector (male)
PROFIBUS DP bus: 9-pin D-connector (female);
power: 9-pin D-connector (male)
Devicenet™ 5-pin M12-connector (male)
FLOW-BUS/Modbus-RTU/ASCII RJ45 modular jack

Technical specifications and dimensions subject to change without notice.

For dimensional drawings and hook-up diagrams please visit the productpage on our website

Downloads

Brochures
μ-FLOW series L01 brochure

Manuals
General manual digital instruments
Operating instructions digital instruments

Hook-up diagrams
Analog IO - RS232
DeviceNet
FLOW-BUS
Modbus
PROFIBUS DP


Bronkhorst High-Tech B.V.

Nijverheidsstraat 1A
NL-7261 AK Ruurlo (NL)
Tel. +31 573 45 88 00
info@bronkhorst.com

Copyright © 2018 Bronkhorst. All rights reserved.     Sitemap     Disclaimer     Privacy note