Bronkhorst

LOW-ΔP-FLOW F-202EV

Mass Flow Controller for low pressure drop or corrosive gas service

  • Large bore capillary (thermal bypass sensor)
  • Very low pressure drop
  • Less sensitivity to humidity or dirt
  • Suitable of corrosive gases
  • Easy to purge
Get a quote Request a demo Support
LOW-ΔP-FLOWF-202EV

Thermal Mass Flow Controllers for low pressure drop or corrosive gas applications

Bronkhorst® model F-202EV Mass Flow Controllers (MFCs) are suited for precise measurement of flow ranges between 0,17…8,5 ln/min and 1…50 ln/min (N2-equivalent). The instruments are particularly suited for corrosive gases or applications with very low differential pressure (∆P). Compared to conventional instruments, LOW-∆P-FLOW MFCs have larger flow channels to minimize the risk of clogging, facilitate cleaning and purging, and cause lower pressure drop (the sensor only requires 0,5 to 5 mbar).

The integrated digital pc-board provides signal and fieldbus conversion as well as PID controller functionality for mass flow control by means of the integrated control valve. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide DeviceNet™, PROFIBUS DP, Modbus, EtherCAT®, PROFINET or FLOW-BUS protocols.


Technical specifications

Measurement / control system
Accuracy (incl. linearity)
(based on actual calibration)
±1%FS
Turndown 1:50 (2...100%)
Multiple fluid capability storage of max. 8 calibration curves
Repeatability < 0,2% Rd
Settling time (controller) 2...3 seconds
Control stability < ±0,1% FS (typical)
Max. Kv-value 6,6 x 10-2
Operating temperature -10…+70°C
Temperature sensitivity < 0,1% FS/°C
Pressure sensitivity 0,1% Rd/bar typical N2
Leak integrity, outboard tested < 2 x 10-9 mbar l/s He
Mounting position horizontal
Warm-up time 30 min. for optimum accuracy
2 min for accuracy ± 2% FS
Mechanical parts
Material (wetted parts) stainless steel 316L or comparable;
other on request
Max. operating pressure 10 bar
Process connections compression type or face seal couplings
Seals standard: Viton®; options: EPDM, Kalrez® (FFKM)
Weight 0,8 kg
Ingress protection (housing) IP40
Electrical properties
Power supply +15…24 Vdc
Max. power consumption Supply
15 V
24 V
at voltage I/O
290 mA
200 mA
at current I/O
320 mA
215 mA
PROFIBUS DP add 53 mA (15 V supply) or 30 mA (24 V supply)
PROFINET add 77 mA (15 V supply) or 48 mA (24 V supply)
EtherCAT® add 66 mA (15 V supply) or 41 mA (24 V supply)
DeviceNet™ add 48 mA (24 V supply)
Analog output/command 0...5 (10) Vdc or 0 (4)…20 mA (sourcing output)
Digital communication standard: RS232;
options: PROFIBUS DP, DeviceNet™, PROFINET, EtherCAT®, Modbus RTU or ASCII, FLOW-BUS
Electrical connection
Analog/RS232 9-pin D-connector (male);
PROFIBUS DP bus: 9-pin D-connector (female);
power: 9-pin D-connector (male);
Devicenet™ 5-pin M12-connector (male);
EtherCAT®/ PROFINET 2 x RJ45 modular jack (in/out)
FLOW-BUS/Modbus-RTU/ASCII RJ45 modular jack

Technical specifications and dimensions subject to change without notice.

For dimensional drawings and hook-up diagrams please visit the productpage on our website

Downloads

Brochures
LOW-ΔP-FLOW Brochure

Manuals
General manual digital instruments
Operating instructions digital instruments

Hook-up diagrams
Analog IO - RS232
DeviceNet
EtherCAT
FLOW-BUS
Modbus
Optional Bus and IO Configurations
PROFIBUS DP
PROFINET

Dimensional drawing


Bronkhorst High-Tech B.V.

Nijverheidsstraat 1A
NL-7261 AK Ruurlo (NL)
Tel. +31 573 45 88 00
info@bronkhorst.com

Copyright © 2018 Bronkhorst. All rights reserved.     Sitemap     Disclaimer     Privacy note